Research Article Details

Article ID: A45167
PMID: 25845713
Source: Magn Reson Med
Title: Sensitivity of chemical shift-encoded fat quantification to calibration of fat MR spectrum.
Abstract: PURPOSE: To evaluate the impact of different fat spectral models on proton density fat fraction quantification using chemical shift-encoded MRI (CSE-MRI). METHODS: In a simulation study, spectral models of fat were compared pairwise. Comparison of magnitude fitting and mixed fitting was performed over a range of echo times and fat fractions. In vivo acquisitions from 41 patients were reconstructed using seven published spectral models of fat. T2-corrected stimulated echo acquisition mode MR spectroscopy was used as a reference. RESULTS: The simulations demonstrated that imperfectly calibrated spectral models of fat result in biases that depend on echo times and fat fraction. Mixed fitting was more robust against this bias than magnitude fitting. Multipeak spectral models showed much smaller differences among themselves than from the single-peak spectral model. In vivo studies showed that all multipeak models agreed better (for mixed fitting, the slope ranged from 0.967 to 1.045 using linear regression) with the reference standard than the single-peak model (for mixed fitting, slope = 0.76). CONCLUSION: It is essential to use a multipeak fat model for accurate quantification of fat with CSE-MRI. Furthermore, fat quantification techniques using multipeak fat models are comparable, and no specific choice of spectral model has been shown to be superior to the rest.
DOI: 10.1002/mrm.25681